Preview

NNC RK Bulletin

Advanced search

STUDY OF DOSIMETRY POSSIBILITIES USING ELECTRON PARAMAGNETIC RESONANCE ON ALANINE FOR ILU-10 ACCELERATOR INP RK

https://doi.org/10.52676/1729-7885-2022-1-36-42

Abstract

Experimental studying possibilities of the electron paramagnetic resonance (EPR) method application for the dose control using alanine deteectors in the process of technological irradiations with brake radiation at the accelerator ILU-10 INP RK has been carried out. The EPR signal stability and sensitivity of original alanine detectors have been studied, serviceability in standard modes was shown in the range of small doses of brake irradiation (0,005–1 kGy); for dose estimation the calibration on gamma-irrdiation was used. The obtained data confirmed the promise of alanine EPR dosimetry at the ILU-10 accelerator INP RK.

About the Authors

Zh. T. Mukan
Institute of Nuclear Physics ME RK; L.N. Gumilyov Eurasian National University
Kazakhstan

Almaty, Nur-sultan



T. A. Seredavina
Institute of Nuclear Physics ME RK
Kazakhstan

Almaty



N. S. Sushkova
Institute of Nuclear Physics ME RK
Kazakhstan

Almaty



I. V. Danko
Institute of Nuclear Physics ME RK
Kazakhstan

Almaty



N. V. Glushchenko
Institute of Nuclear Physics ME RK

Almaty



References

1. Alimov, A. S. Prakticheskoe primenenie elektronnykh uskoriteley [BIIYaF im. D. V. Skobel'tsyna] / A. S. Alimov. – http://www.knigi.konflib.ru/8bezopasnost/66691.

2. Delincée, M. H. / Food Irradiation – Chemical Aspects. European School of Advanced Studies on Nuclear and Ionizing Radiation Technologies. // University of Pavia. – April 2005. – 43 p.

3. ISO/ASTM №51607; ASTM Standard E1607 “Standard Practice for Use of the Alanine EPR Dosimetry System”.

4. GOST 31652-2012. Produkty pishchevye. Metod elektronnogo paramagnitnogo rezonansa dlya vyyavleniya radiatsionno-obrabotannykh produktov, soderzhashchikh kristallicheskiy sakhar. – Vved, 2013-07-01. – Moscow: Standartinform, 2013.

5. Pivovarov, S. Some peculiarities and complications of ESR dosimetry in high dose / S. Pivovarov, A. Rukhin, T. Seredavina et al. // Techniques for High Dose Dosimetry. IAEA-TECDOC-1070. – 1999. – P. 221–226.

6. Pivovarov, S.P. Izuchenie svoystv EPR alanina, obluchennogo neytronami / S.P. Pivovarov, S.V. Zhdanov, T.A. Seredavina, i dr.// Tezisy 3-y mezhd. nauch. konf. “Yadernaya i radiatsionnaya fizika”, 4–7 june 2001, Almaty, Kazakhstan. – Almaty, 2001. – P. 448.

7. Pat. No 13594 RK. Sposob izmereniya dozy neytronnogo oblucheniya i individual'nyy dozimetr /. Pivovarov, S.P., Pavshuk V.A., Kadyrzhanov K.K. i dr., 2011.

8. Seredavina, T.A. Vozmozhnosti EPR-dozimetricheskogo kontrolya radiatsionnoy obrabotki pishchevoy i meditsinskoy produktsii / T.A. Seredavina, A.B. Rukhin, Zh.T. Mukan i dr. // Tezisy VII mezhd. nauchno-praktich. konf. “Semipalatinskiy ispytatel'nyy poligon. Radiatsionnoe nasledie i perspektivy razvitiya”, 21–23 sept. 2016, Kurchatov, Kazakhstan. – Kurchatov, 2016. – P. 85–86.

9. T.Seredavina, O.Stakhov, N.Sushkova, A.Nurkasymova. EPR-investigation of irradiated imported foodstuff and parameters of EPR signals. Eurasian Chemical Technological Journal. 2012. V.14, №4. P.343-349.

10. Seredavina, T.A. Study of Irradiated Food Products and Control by EPR Method in the Kazakhstan Republic / T.A Seredavina, N.S. Sushkov, A.B. Rukhin, Zh.T. Mukan // Proceedings of 3rd International Conference on Food and Biosystems Engineering. Rhodes, Greece. – 2017. – PP. 260–263.

11. Rukhin, A.B. Adaptatsiya alaninovoy EPR-dozimetrii k usloviyam tekhnologicheskikh oblucheniy elektronami v IYaF RK / Rukhin, A.B., Seredavina T.A., Sushkova N.S., Mukan Zh.K. // Tezisy dokl. mezhd. nauch. foruma “Yadernaya nauka i tekhnologii”, posvyashch. 60-letiyu Instituta yadernoy fiziki (11-ya mezhd. konf. “Yadernaya i radiatsionnaya fizika”; mezhd. konf. “Yadro-2017” (67-e Soveshchanie po yadernoy spektroskopii i strukture atomnogo yadra); 8-ya Evraziyskaya konf. “Yadernaya nauka i ee primenenie”), 12–15 sentyabrya 2017 g., Almaty, Kazakhstan / [Otv. red. E. A. Kenzhin]. – Almaty: RGP IYaF, 2017. – P. 329–330.

12. ISO/ASTM 51261 - Standard Guide for Selection and Calibration of Dosimetry Systems for Radiation Processing.

13. Ivannikov A. Signaly v alanine pri alpha- i gamma-obluchenii. Radiatsiya i risk. – Obninsk. – 2016. – Vol. 25, – No. 1. – P. 85–93.

14. Institute of Nuclear Chemistry and Technology, Warsaw, Poland. Official web-site – http://www.ichtj.waw.pl/ichtj/


Review

For citations:


Mukan Zh.T., Seredavina T.A., Sushkova N.S., Danko I.V., Glushchenko N.V. STUDY OF DOSIMETRY POSSIBILITIES USING ELECTRON PARAMAGNETIC RESONANCE ON ALANINE FOR ILU-10 ACCELERATOR INP RK. NNC RK Bulletin. 2022;(1):36-42. (In Russ.) https://doi.org/10.52676/1729-7885-2022-1-36-42

Views: 476


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 1729-7516 (Print)
ISSN 1729-7885 (Online)